18.03.2013 Measurement of the luminescence lifetime with the nanosecond range using the Fluorolog-3 spectrofluorimeter

Published Date:

The luminescence lifetime of the fluorescein solution was measured using the Fluorolog-3spectrofluorimeter. . The luminescence excitation of the sample was generated by the LED with 495 nm wavelength and with the pulse duration less than 1.5 ns. The luminescence was obtained at the wavelength λem=517 nm. Received luminescence lifetime is τ=3.96±0.01 ns (table lifetime of the fluorescein is 4.0 ns).

nwes6e

Luminescence kinetics of the fluorescein solution (λem=517 nm)